Cleanroom classification is a subject anyone working in pharmaceutical manufacturing knows in outline: the ISO classes, the GMP grades, the particle concentration limits. Far less discussed is the metrological aspect of the issue, namely that a class is not a construction characteristic of the room but the result of a measurement.
And if it is the result of a measurement, it depends entirely on the instruments used to take it. A correctly designed room, verified with uncalibrated instrumentation, is not a classified cleanroom: it is a presumed one.
How cleanrooms are classified
The international reference is ISO 14644-1, which defines nine air cleanliness classes based on the concentration of airborne particles per cubic metre, ranging from ISO 1 (the cleanest) to ISO 9. In pharmaceutical manufacturing this sits alongside the grade classification set out in EU GMP Annex 1, which identifies four grades by letter, from A to D.
The two systems do not coincide but they map onto each other. This is the essential scheme:
| GMP grade | ISO class | Typical use |
| A | ISO 5 (at rest and in operation) | Critical aseptic operations, sterile filling |
| B | ISO 5 at rest, ISO 7 in operation | Background environment directly supporting grade A |
| C | ISO 7 at rest, ISO 8 in operation | Less critical stages of sterile manufacturing |
| D | ISO 8 at rest | Preparation and upstream processing |
The difference between the at-rest and in-operation figures leads us to the second key concept: the occupancy state.
Occupancy states: why the same room has different classes
ISO 14644-1 provides for three states in which classification may be declared. As-built, with the installation complete and functioning but no equipment or personnel present. At rest, with the equipment installed and running but no personnel present. In operation, with production under way and personnel present.
A classification that does not state the occupancy state it refers to is incomplete and cannot be used for compliance purposes. It is also why the instruments used have to perform reliably under markedly different conditions.
The measuring instruments required for classification
Classification and the resulting monitoring involve several families of instruments, each dedicated to a different parameter.
Optical particle counters
These are the instruments that determine the class: they measure the concentration of airborne particles at the threshold sizes set by the standard. Their calibration follows ISO 21501-4 and covers particle size, counting efficiency and sample flow rate. An error in the flow rate translates directly into an error in the calculated concentration, and therefore in the class assigned to the room.
Microbiological air samplers
In aseptic areas particle counting is not enough: Annex 1 also requires monitoring of microbiological contamination, carried out with active samplers that draw a known volume of air onto culture medium. The critical parameter to calibrate is the sampling flow rate: if the volume actually drawn differs from the nominal one, the result expressed in colony-forming units per cubic metre is wrong in proportion.
Air velocity and flow rate instruments
Anemometers and balometers verify the unidirectional airflow velocity in grade A zones and the air change rates in turbulently ventilated areas. These measurements determine the room’s capacity to remove contamination generated during operations, and the instruments must be calibrated across the entire range actually in use, not merely at convenient points.
Differential pressure gauges
The pressure cascades between rooms of different classification are the mechanism that prevents contamination from spreading to the most critical areas. The differential pressure gauges that monitor them work on very small differences, where the instrument’s uncertainty weighs heavily on the compliance decision.
Temperature and humidity probes
Temperature and relative humidity do not define the class, but are controlled environmental parameters: they affect operator comfort, particle generation and product stability. They are therefore included in the calibration programme alongside the other instruments.
Why calibration underpins classification
Classification is not a permanent result. ISO 14644-2 and Annex 1 require it to be reverified at defined intervals, more frequently for the more critical grades, and require the room to be under continuous monitoring between one verification and the next. Each time, the data produced is worth exactly what the instrument that generated it is worth.
During an inspection this translates into a very concrete request: the calibration certificates of the instruments used in classification and monitoring, with documented metrological traceability and stated expanded uncertainty. The same requirements apply as those already covered in relation to calibration of instruments under GMP, here in a context where the measured parameter directly defines the compliance of the room.
Checklist for instrument compliance
- The declared classification always states the occupancy state it refers to
- Optical particle counters are calibrated to ISO 21501-4
- The sampling flow rate of microbiological samplers is verified and documented
- The calibration range of each instrument covers its actual range of use in the room
- Certificates state the expanded uncertainty and traceability to primary standards
- The issuing laboratory is accredited to ISO/IEC 17025 and recognised under the ILAC-MRA arrangements
To check that the instruments used to classify and monitor your cleanrooms are calibrated appropriately for GMP requirements, contact the Altiorlab team: we will review your instruments and their certificates with you.